TU810 In-line Turbidity Sensor

Turbidity sensor for continuous measurement in pipes, with nephelometric method ISO 7027 and selectable ranges of 0–4, 0–40, 0–400 and 0–4000 NTU.

The SB&C Electronics TU810 in-line turbidity sensor It is a high-precision optical sensor designed for the continuous turbidity measurement in water and process liquids. Its nephelometric measurement technology complies with the standards ISO 7027 and EN 27027, It uses an 890 nm infrared LED to obtain stable measurements, minimizing the influence of sample color and ambient light.

This turbidity sensor It incorporates a built-in preamplifier and is designed to work in conjunction with the controllers. TU7685 and TU7685.010, allowing measurement from very low turbidity waters to processes with high concentrations, thanks to its different measurement scales up to 4000 NTU.

Its chemical-resistant PVC body, IP67 connector protection, and the ability to install cables up to 150 meters long allow its use in drinking water treatment plants, filtration systems, swimming pools, process water, mineral water, ultrafiltration, sedimentation, and water quality control in industrial applications.

To obtain maximum accuracy, especially on the scale of 0–4 NTU, The manufacturer recommends installing the sensor in the TU910 flow cell or in a PVC T that can be supplied. These accessories maintain a constant flow rate of the sample, reduce the appearance of air bubbles in the optical zone, and facilitate both cleaning and calibration using formazin standards.

Installation of the TU810 online turbidimeter with TU910 flow cell
Installation of the TU810 turbidity sensor in line using the PVC flow cell for continuous turbidity measurement.

The set formed by the TU810 sensor, the TU910 flow cell and the TU7685 controller It constitutes a complete solution for continuous turbidity monitoring in industrial processes and water treatment plants, providing a reliable, stable measurement with very little maintenance.

Properties

ReferenceTU810
Product typeAmplified turbidity sensor for online measurement
Measurement methodNephelometric, in accordance with ISO 7027 and EN 27027
Light source890nm infrared LED
Measuring range0–4000 NTU
Scales available with TU7685 controller0–4,000; 0–40,00; 0–400,0 and 0–4000 NTU
Resolution0.001 NTU on the 0–4,000 NTU scale
0.01 NTU on the 0–40.00 NTU scale
0.1 NTU on the 0–400.0 NTU scale
1 NTU on the 0–4000 NTU scale
Precision±5 % of the reading between 0 and 400 NTU
±10 % of the reading between 400 and 4000 NTU
Response time10 seconds
PreamplifierIncorporated
Internal detectionEmpty cell and dirty optical window control
Sensor power supply±12 VDC supplied by the TU7685 controller
Compatible driversB&C Electronics TU7685 and TU7685.010
Ambient temperature0 to 50 °C
Sample temperature0 to 50 °C
Maximum sample pressure6 bar at 20 °C
Body materialPVC
O-ring materialNBR, nitrile rubber
Optical window materialAcrylic
Sensor diameter40 mm
Connector protectionIP67
Maximum cable length150 m
Available cables10 or 30 m with connector
Direct pipe mounting2-inch T-connection, DN50
Low turbidity accessoryTU910 flow cell recommended for the 0–4 NTU scale
TU810 B&C Electronics Inline Turbidimeter
Turbidimeter TU810 TU910 B&C Electronics

Other products

Resistive linear position sensor with spring probe • up to 200 mm • ±0.1 % FS
Monitran-Logo
ISO 10816-3 Standard | Color TFT Display | 100-Record Memory | Bearing Analysis | USB/CSV Export | >18h Li-ion Battery | Audio/Headphone Output | Low-Pass Filters | RMS and Peak Measurement | Industrial Rubber Housing
Anemometer with pulse output, proportional to wind speed, industrial use
Official logo of ELCIS Encoder, Italian manufacturer of incremental and absolute encoders and industrial measurement systems.
Ø59.5 mm | Hollow shaft Ø4 to Ø10 mm | Elastic coupling | Rotating body for aligning the electrical zero with the mechanical zero | Max. frequency 60 kHz (100 kHz optional) | Resolution 1 to 2540 imp/v | Anodized aluminum base and cover | Weight 110 g | IP53 on the shaft side | NPN, PNP, push-pull, TTL and line driver outputs
Official logo of ELCIS Encoder, Italian manufacturer of incremental and absolute encoders and industrial measurement systems.
Resolution up to 65536 steps/revolution (16-bit) | Up to 4096 revolutions (12-bit) | SSI interface with preset · Gray or Binary | Sine/Cosine output 1024 ppr or RS422 up to 16384 ppr | Power supply 10–30 V · max. consumption 200 mA | Hollow spindle H7: Ø6, Ø8, Ø10, Ø12 or Ø15 mm | Anti-rotation pin | Spindle-side protection IP64 (IP66 sealed) | Max. continuous speed 6000 rpm (3000 rpm sealed) | Operating temperature -20…+85 °C
Official logo of Novotechnik, manufacturer of linear sensors, position transducers, and industrial encoders
Potentiometric linear displacement sensor with articulated mounting, travels from 10 to 300 mm and IP54
Radar level sensor, range up to 10 meters.
THERMOHYGROMETER H. Rel., Dew point, Temp., Wet bulb
Official logo of ELCIS Encoder, Italian manufacturer of incremental and absolute encoders and industrial measurement systems.
ATEX Ex d IIC T6 Gb certified | B10 flange (Ø115) with Ø85 h7 centering | Aluminum housing | Single-turn 7 to 12-bit | Up to 4096 steps per revolution | Ø10 h7 stainless steel shaft | Parallel, SSI serial (RS422) or analog output | IP65 standard and IP66 on the shaft with Z option | Max. radial/axial load 10 kg | Power supply 5 V, 15 V, 8–24 V or 18–28 V
Shopping Basket