SCO2 Carbon Dioxide Sensor
- Manufacturer: Nesa
- Category: Gases - Particles, Meteorology – Climatology, SENSORS, Gas Sensors
The NESA SCO2 is a carbon dioxide sensor for monitoring air quality in environments with crowds and non-continuous flow: art galleries, museums, caves, auditoriums, schools, hospitals, greenhouses and animal breeding.
Dual-wavelength NDIR measurement system, typical measurement range 0–2000 ppm or 0–5000 ppm and accuracy ±(20 ppm + 2% of the measurement). Output 4–20 mA or RS485/Modbus. Each instrument is supplied with a calibration certificate.
Key points
- Typical measurement range 0÷2000ppm or 0÷5000ppm; accuracy ±(20ppm+2% of measurement)
- Response time <120s, stabilization time <15s; long-term stability 5% of field / 5 years
- Salida 4÷20mA (RL<500 Ohm) o RS485 / ModBus RTU; salidas del borne CO2, Temp y Alarmaa
- Power supply 12÷30Vdc; consumption <2W @ 24Vdc
- Protections: against short circuits and overvoltage; polarity reversal and transients
- Operating temperature -25 ÷ +60°C, IP21, indoor enclosure; ABS and stainless steel screws; 120g; CE
Technical Frequently Asked Questions
What is its measurement range and accuracy?
Typical measurement range 0÷2000ppm or 0÷5000ppm, with accuracy ±(20ppm+2% of the measurement).
What exits does it have?
4÷20mA (RL<500 Ohm) o RS485 / ModBus RTU, además de las salidas del borne CO2, Temp y Alarmaa.
How do you calibrate it?
Cada instrumento se calibra y verifica por comparación con un instrumento patrón certificado como patrón primario de la WMO y se entrega con certificado de calibración.
Properties
| Feature | Specification |
|---|---|
| Measurement and operating principle | |
| Transducer type | Dual wavelength non-infrared (NDIR) technology |
| Measurement principle | NDIR (Non-Dispersive Infrared Technology) infrared sensor which, thanks to the use of a double filter and a particular measurement technique, compensates for the effect of its aging |
| Typical measurement range | 0-2000ppm or 0-5000ppm |
| CO2 accuracy (fs 0÷2000ppm) | ±(20ppm+2% of the measurement) |
| CO2 accuracy (fs 0÷5000ppm) | ±(20ppm+2% of the measurement) |
| Long-term stability | 5% of the field / 5 years |
| Response time | <120s |
| Stabilization time | <15s |
| Calibration | Cada instrumento se calibra y verifica por comparación con un instrumento patrón certificado como patrón primario de la WMO y se entrega con certificado de calibración |
| Exit signals | |
| Output signal 4÷20mA | RL<500 Ohm |
| RS485 output signal | RS485 / ModBus RTU |
| Terminal exits | CO2, Temp, Alarmaa |
| Power supply, environment, and mechanical data | |
| Feeding | 12÷30Vdc |
| Consumption | <2W @ 24Vdc |
| Protections | Protection against short circuits and overvoltage; polarity reversal and transients |
| Operating temperature | -25 ÷ +60°C (0÷100% RH, non-condensing) |
| Electronics protection | IP21 |
| Made in | ABS and stainless steel screws |
| Dimensions | View map |
| Weight | 120g |
| Regulatory compliance | EC |
| Typical applications | |
| Indoor air quality control | Environments with crowds and non-continuous flow of people: art galleries, museums, caves, auditoriums, schools, hospitals, greenhouses, animal breeding |