RSG2std Class A Pyranometer
- Manufacturer: Nesa
- Category: Meteorology – Climatology, Solar radiation - Light, SENSORS
El NESA RSG2std es un piranómetro de clase A según ISO 9060 (patrón secundario) de doble termopila y alta velocidad de respuesta, para la medida de la radiación solar global entre 0,3 y 3 µm con respuesta en coseno. Conforme a WMO para la monitorización ambiental y apto para aplicaciones fotovoltaicas (ISO 9060 e ISO/IEC 17025).
Measurement range 0 to 2000 W/m², zero offset < 7 W/m², response time < 10 s, nonlinearity < ±0.5%, and expected daily uncertainty < 2%. Double quartz glass dome (K5), IP67 aluminum housing, and native output in µV, 0–1 Vdc, 4–20 mA, or RS485/Modbus. Supplied with a 5 m cable and calibration report.
Key points
- clase A (ISO 9060, patrón secundario); conforme a WMO
- 0 ÷ 2000 W/m², 0.3 ÷ 3 µm, 10 µV/(W/m²)
- Zero offset < 7 W/m², stability < ±0.8% per year, non-linearity < ±0.5%
- Outputs µV, 0÷1Vdc, 4÷20mA or RS485 / Modbus (ID 003)
- Anodized aluminum, IP67, -40 ÷ +80 °C, < 850 g
Technical Frequently Asked Questions
What outputs does the RSG2std have?
RSG2Std-N: natural thermopile, 10 µV/(W/m²). RSG2Std-A: 0÷1Vdc. RSG2Std-B: 4÷20mA. RSG2Std-c: RS485 / Modbus with optional MCS module (standard ID 003).
What makes it stand out?
Es de clase A (patrón secundario): doble termopila, resolución < 2 W/m², respuesta < 10 s, offset de cero < 7 W/m², estabilidad < ±0,8 % al año, respuesta al coseno < ±10 W/m² e incertidumbre diaria prevista < 2 %.
How do I order it?
Sensor (RSG2Std), output (A, B, N or C), sensor-datalogger cable (05, 10 or xx, length in meters to be specified) and bracket (SS1 of l = 500 mm or SS3 of l = 900 mm). Manufacturer's example: RSG2Std N 10 SS1.
Properties
| Feature | Specification |
|---|---|
| Measurement and operating principle | |
| Measured magnitude | Radiación solar global (respuesta en coseno) |
| Measurement range | 0 ÷ 2000 W/m² |
| Spectral range | 0.3 µm ÷ 3 µm (AM1.5G) |
| Typical sensitivity | 10 µV/(W/m²) over 2π sr |
| Typical output constant | 10 mV/(kW/m²) |
| Resolution | < 2 W/m² |
| Response time | < 10 s |
| Zero offset | < 7 W/m² (at 200 W/m²); < ±2 W/m² (ΔT = 5 K/h) |
| Long-term stability (1 year) | < ±0.8% |
| Cosine response | < ±10 W/m² |
| Temperature response (ΔT 50 K) | < 2% |
| Nonlinearity | < ±0.5 % |
| Response to tilt (0 – 90°) | < ±0.5 % |
| Expected daily uncertainty | < 2% |
| Transducer | Double thermopile |
| Classification | Clase A según ISO 9060 (patrón secundario) |
| Regulatory compliance | ISO 9060; WMO; CE |
| Output signals and interface | |
| Output (version N) | 10 µV/(W/m²), direct from thermopile |
| Output (version A) | 0÷1Vdc |
| Output (version B) | 4÷20mA |
| Output (version C) | RS485 / Modbus, with optional MCS module; Standard ID 003 |
| Output impedance | < 900 Ω |
| Calibration | By comparison with a primary standard instrument of superior class and known constant |
| Power supply, environment, and mechanical data | |
| Food and consumption | Self-powered in the N version; 10 ÷ 30Vdc < 0.1 W in the others |
| Protections | Protection against short circuits and lightning strikes |
| Operating conditions | -40 to +80 °C |
| Degree of protection | IP67 |
| Material | Anodized aluminum and stainless steel screws |
| Dimensions | 165 x 105 mm (see drawing) |
| Weight | < 850 g |
| Reference | Configuration |
|---|---|
| RSG2Std-ARSG2StdA | Model: RSG2Std · Output signal: 0÷1Vdc · Supply: With 5m cable calibration report |
| RSG2Std-BRSG2StdB | Model: RSG2Std · Output signal: 4÷20mA · Supply: With 5m cable calibration report |
| RSG2Std-NRSG2StdN | Model: RSG2Std · Output signal: Natural in µV · Supply: With 5m cable calibration report |
| RSG2Std-cRSG2Stdc | Model: RSG2Std · Output signal: RS485 / Modbus (Standard ID 003) · Supply: With 5m cable calibration report |