e-Bar internal barometer option for Evolution datalogger
- Manufacturer: Nesa
- Category: Meteorology – Climatology, Pressure, Atmospheric pressure, SENSORS
The NESA e-Bar is an internal barometer (atmospheric pressure sensor) option for the Evolution datalogger, featuring a piezoresistive bridge transducer. Typical measurement range is 500–1100 hPa (programmable), and accuracy is <0.4 hPa @ 20°C.
Because it's mounted on the datalogger's internal electronics, it doesn't use the datalogger's physical channels. It's only available as an option when ordered with the Evolution datalogger.
Key points
- Typical measurement range 500 ÷ 1100 hPa (programmable), sensitivity 0.05 hPa
- Resolution <0.15 hPa @ 20°C, accuracy <0.4 hPa @ 20°C, calibration uncertainty <0.15 hPa, response time < 2 s
- Automatic temperature compensation -10°C ÷ +70°C; I2CBus or SDI12 interface directly to the datalogger's µP
- Output signal linearly proportional to the measured atmospheric pressure
- Range and correction configurable from the datalogger
- 10.5÷24Vdc, <0.01W and <0.5 mA, -40 ÷ +70°C
- It is only activated at the factory (requires calibration) and during the manufacturing of the datalogger
- Compliant with CE standards
Technical Frequently Asked Questions
Can the e-Bar be ordered separately?
No. It is only offered as an option in conjunction with the Evolution datalogger.
Does it use up datalogger channels?
No. Because it is mounted on the internal electronics of the datalogger, it does not use the datalogger's physical channels, which maintain their ability to connect to external sensors.
What is its fuel consumption?
The technical data table gives 10.5–24 Vdc and <0.01 W, and the manufacturer's text states <0.5 mA. This does not alter the datalogger's power consumption.
Can it be activated after the datalogger has been manufactured?
No. It is only activated at the factory (requires calibration) and during the manufacturing of the datalogger.
What interface does it use?
I2CBus or SDI12 directly to the datalogger's µP.
What can be configured?
The range and correction, from the datalogger.
What applications is it suitable for?
Meteorology; tunnels and highways; climatology; sports and winter resorts; construction site safety; industrial buildings.
How do I order it?
Sensor: Evolution; accessory: e-Bar. Manufacturer example: Evolution e-bar.
Properties
| Feature | Specification |
|---|---|
| Measurement and operating principle | |
| Measured magnitude | Atmospheric pressure |
| Typical measurement range | 500 ÷ 1100 hPa (programmable) |
| Measurement principle | piezoresistive bridge transducer |
| Transducer type | Piezoresistive |
| Sensitivity | 0.05 hPa |
| Resolution | <0.15 hPa @ 20°C |
| Precision | <0.4 hPa @ 20°C |
| Calibration uncertainty | <0.15 hPa |
| Response time | < 2 s |
| Temperature response | Self-compensation -10°C ÷ +70°C |
| Regulatory compliance | EC |
| Output signals and interface | |
| Exit sign | Linearly proportional to the measured atmospheric pressure |
| Interface signal | I2CBus or SDI12 directly to the datalogger's µP |
| Configurable parameters | Range and correction, from the datalogger |
| Datalogger channels | It does not use physical datalogger channels. |
| Integration | Activatable option on the Evolution datalogger, mounted on its internal electronics |
| Analog inputs of the datalogger | It does not reduce their number, as it is an integrated device. |
| Activation | Only at the factory (requires calibration) and during the manufacturing of the datalogger; it is not possible to activate it afterward |
| Power supply, environment and calibration | |
| Food and consumption | 10.5÷24Vdc, <0.01W |
| Current consumption | <0.5 mA |
| Operating conditions | -40 to +70°C |
| Work environment | Inside the datalogger (controlled environment) |
| Envelope | Inside the ventilated datalogger box |
| Calibration | Each instrument is calibrated and verified with a pressure generator/regulator connected to a primary gauge |
| Typical applications | |
| Application | Meteorology |
| Application | Tunnels, highways |
| Application | Climatology |
| Application | Sports and winter resorts |
| Application | Safety at construction sites |
| Application | Industrial buildings |
| Reference | Configuration |
|---|---|
| e-BareBar | Supply: Electronic barometer inside the Evolution datalogger; programmable range, standard 500-1100 hPa; read directly by the datalogger as it is mounted on the motherboard |