KELLER ECO2 Digital Pressure Gauge
He KELLER ECO2 digital pressure gauge It is a compact and economical instrument for measuring and testing pressure in industrial installations. It is available with ranges of 0…31 absolute bar and 0…300 bar, It offers accuracy of up to ±0.5 % of the full scale at room temperature and does not require an external power supply, as it operates using a replaceable CR2430 lithium battery.
Its two-line LCD screen simultaneously displays the current pressure and the last recorded maximum value. Pressing the ENTER button also displays, for five seconds, the minimum value measured since the last reset. This function facilitates checking pressure variations, extreme values, and potential drops or increases during maintenance and monitoring operations.
Compact digital pressure gauge for industrial applications
The KELLER ECO2 is a simple digital alternative to conventional mechanical pressure gauges when a clear reading, different pressure units, and higher resolution than an analog dial are required. The pressure is updated at a frequency of two measurements per second and is displayed on an LCD screen with two rows of 4.5 digits.
The instrument incorporates an isolated and encapsulated piezoresistive pressure sensor. The pressure connection and the sensor's diaphragm are made of AISI 316L stainless steel. Its housing is protected. IP65, suitable against the entry of dust and water jets.
Available ranges of the KELLER ECO2
The ECO2 digital pressure gauge is supplied in two versions:
- ECO2 of 31 bar: range of 0…31 bar absolute, resolution of 0.01 bar and overload resistance up to 60 bar.
- ECO2 300 bar: range of 0…300 bar, resolution of 0.1 bar and overload resistance up to 400 bar.
The 31 bar version comes factory-set to zero at a vacuum, i.e., 0 bar absolute. When a measurement relative to atmospheric pressure is desired, the function can be used. ZERO SET with the pressure gauge exposed to ambient pressure.
The 300 bar version uses 1 bar absolute as its zero reference. Before selecting the range, it is advisable to consider the normal working pressure, potential overpressures, and the required resolution.
Accuracy and temperature performance
The specified accuracy at an ambient temperature of 20 to 25 °C is ≤ ±0.5 % FS. This specification includes nonlinearity, pressure hysteresis, lack of repeatability, and zero-point and amplification deviations.
Over the entire compensated range of 0 to 50 °C, the total error band is ≤ ±1.0 % FS. This band represents the maximum possible deviation within the specified pressure and temperature ranges, making it the most suitable data for assessing the maximum error under real operating conditions.
Long-term stability is up to ±0.2 % FS per year under reference conditions. When the pressure gauge is used in quality control, certified maintenance, or processes requiring traceability, KELLER recommends annual recalibration.
Current pressure reading and minimum and maximum values
The upper display continuously shows the current pressure. The lower display normally shows the maximum value recorded since the last reset. Briefly pressing ENTER will display the minimum stored value for five seconds.
The function RESET It sets the minimum and maximum values to the current pressure. This allows a new check to be started without needing to turn off or remove the pressure gauge.
When the applied pressure exceeds the display range, the screen shows an overflow or underflow warning. If the pressure exceeds the instrument's measurement range, the last valid value flashes as an overload warning.
Functions of the ECO2 pressure gauge
- RESET: resets the stored minimum and maximum values.
- ZERO SET: sets the applied pressure as a new zero reference.
- ZERO RESET: restores the zero setting made at the factory.
- MIN/MAX: It allows you to view the minimum and maximum measured values.
- OFF: Manual instrument shutdown.
- CONT ON: disables automatic shutdown for continuous operation.
- CONT OFF: It activates automatic shut-off after approximately 15 minutes without any pulses.
- BAT LOW: It alerts you on the screen when the battery needs to be replaced.
Selectable pressure units
The user can change the unit of measurement directly using the two front buttons, without the need for software or external accessories. The available units are:
- bar
- amber
- hPa
- kPa
- MPa
- PSI
- kp/cm²
The selected settings are retained after turning off the pressure gauge.
Swivel G1/4 adapter
The ECO2 body incorporates a male connection 7/16-20 UNF. Each instrument is delivered with a swivel adapter type Swivel with G1/4 male connection on the process side.
The Swivel system incorporates O-rings and allows the pressure gauge to be rotated to orient the display in the most comfortable reading position, without the final position depending on the tightening point of a conventional thread. The maximum tightening torque indicated for the adapter is 50 Nm.
Optional G1/8, 1/4-18 NPT and G1/4 adapters compliant with EN 837 without orientation function are also available.
Battery operated
The KELLER ECO2 operates using a lithium coin cell battery. CR2430 3V, factory-installed and included in the delivery. The battery range extends up to 1,000 hours of continuous operation.
To extend its battery life, the instrument can automatically shut off approximately 15 minutes after the last press. This function can be deactivated when continuous readings are required. When the battery is running low, the BAT LOW indicator will appear on the display.
Operating temperatures
- Compensated temperature range: 0…50 °C.
- Medium temperature: −10…60 °C.
- Ambient temperature: 0…50 °C.
- Storage temperature: −20…70 °C.
Outside the 0 to 50 °C range, display readability may be affected. The ambient temperature should not be confused with the storage temperature or the range within which the total error margin is guaranteed.
Common applications
Thanks to its compact size, independent power supply and two-button operation, the KELLER ECO2 digital pressure gauge is suitable for:
- Pressure checks in industrial installations.
- Maintenance and technical service operations.
- Temporary supervision of hydraulic and pneumatic circuits.
- Checking valves, regulators and pressurized equipment.
- Pressure control in machines and auxiliary equipment.
- Supervision of glass bottle filling processes.
- Replacement of analog pressure gauges with a digital display.
- Detection of minimum and maximum values during a test.
Contents of the supply
- KELLER ECO2 digital pressure gauge.
- Renata CR2430 lithium battery installed.
- Swivel G1/4 male adjustable adapter.
- Copper gasket for G1/4 connection.
- Instruction manual.
Available references
- 303060.0001: KELLER ECO2, range of 0…31 bar absolute.
- 303060.0002: KELLER ECO2, range 0…300 bar.
Optional accessories include a protective rubber cover, a carrying bag, and KELLER calibration certificates with 5 or 11 measurement points.
ECO2-Ei version for areas with risk of explosion
Standard ECO2 should not be used in potentially explosive atmospheres. For these applications, the following is available: KELLER ECO2-Ei with intrinsic safety, Available in 31 and 300 bar versions and with ATEX and IECEx certification. The ECO2-Ei version must be specifically requested.
Selection and supply of the KELLER ECO2
GUEMISA supplies the KELLER ECO2 digital pressure gauge and can advise on range selection, process connection, accessories, and calibration certificate. To prepare a quote, please specify the maximum working pressure, fluid type, required connection, and whether the installation is located in an explosion-proof zone.
Properties
| Model and operation | |
|---|---|
| Manufacturer | KELLER Pressure |
| Model | ECO2 |
| Guy | Compact battery-powered digital pressure gauge |
| Measurement principle | Piezoresistive pressure sensor, isolated and encapsulated |
| Communication interface | None |
| Data logging | It does not have a recorder; it only memorizes the minimum and maximum values since the last RESET |
| Pressure ranges | |
| ECO2 range / 31 bar | 0…31 bar absolute, PAA |
| Zero adjustment, version 31 bar | 0 bar absolute, empty; a relative zero can be set using ZERO SET |
| ECO2 range / 300 bar | 0…300 bar, PA |
| Zero adjustment, 300 bar version | Factory reference at 1 bar absolute |
| Maximum overload, 31 bar version | 60 bar |
| Maximum overload, 300 bar version | 400 bar |
| Measurement range reservation | ≥ ±10 % with display of valid values outside the nominal range |
| Accuracy and performance | |
| Accuracy at 20…25 °C | ≤ ±0.5 % FS |
| Parameters included in accuracy | BFSL nonlinearity, hysteresis, lack of repeatability, zero deviations, and amplification |
| Total error band at 0…50 °C | ≤ ±1.0 % FS |
| Long-term stability | ≤ ±0.2 % FS per year, under reference conditions |
| Measurement frequency | 2 measurements per second |
| Resolution, range of 31 bar | 0.01 bar |
| Resolution, range of 300 bar | 0.1 bar |
| Recommended recalibration interval | Annually when the application requires maintaining traceability and specified performance |
| Screen | |
| Screen type | two-line LCD |
| Ability | Two rows of 4.5 digits |
| LCD dimensions | 27.8 × 30.0 mm |
| Height of the upper digits | 8 mm |
| Height of lower digits | 7 mm |
| Upper indication | Current pressure |
| Lower indication | Maximum or minimum value measured since the last RESET |
| Functions | |
| MIN/MAX | Display of the minimum and maximum recorded values |
| RESET | Set the minimum and maximum values to the current pressure. |
| ZERO SET | Set the applied pressure as the new zero reference. |
| ZERO RESET | Restores the factory-set zero point |
| Automatic shut-off | After approximately 15 minutes without a pulse |
| Continuous operation | Yes, using the CONT ON function. |
| Low battery warning | BAT LOW Indication |
| Off-screen notice | OFL for overflow and UFL for underflow |
| Overload warning | The last valid value flashes when the measurement range is exceeded. |
| Pressure units | |
| Configurable units | bar, mbar, hPa, kPa, MPa, PSI and kp/cm² |
| Temperature | |
| Compensated range | 0…50 °C |
| Medium temperature | -10…60 °C |
| Ambient temperature | 0…50 °C |
| Storage temperature | -20…70 °C |
| Feeding | |
| Battery | CR2430 lithium, 3V |
| Battery included | Yes, Renata CR2430 installed |
| Autonomy | Up to 1,000 hours of continuous operation |
| Pressure connection | |
| Instrument connection | 7/16-20 UNF male |
| Connection to the process included | G1/4 Male Swivel Adapter |
| Orientation | Rotating adapter for orienting the screen in the reading position |
| Maximum adapter torque | 50 Nm |
| Optional adapters | G1/8, 1/4-18 NPT and G1/4 EN 837 without swivel function |
| Materials | |
| Pressure connection | AISI 316L stainless steel |
| Separating membrane | AISI 316L stainless steel |
| Internal transducer gasket | Without an internal board |
| Internal transmission fluid | Silicone oil |
| Swivel adapter insert | Nickel-plated brass |
| Swivel Joints | NBR |
| Adapter | Galvanized steel |
| adapter outer gasket | NBR |
| Screen housing | Polyamide |
| Dimensions and protection | |
| Dimensions without case | 59 × 95 × 32 mm |
| Dimensions with protective case | 70 × 101 × 42 mm |
| Approximate weight | 125g, without case or adapter |
| Degree of protection | IP65 |
| Accordance | |
| Electromagnetic compatibility | Directive 2014/30/EU |
| CEM Standards | EN IEC 61326-1, EN IEC 61326-2-3 and applicable EN IEC 61000-6 series |
| Protection against explosive atmospheres | Not in the standard ECO2; select the ECO2-Ei model |
| Order references | |
| ECO2 / 31 bar | 303060.0001 |
| ECO2 / 300 bar | 303060.0002 |
| Supplies and accessories | |
| Standard content | Pressure gauge, CR2430 battery, G1/4 swivel adapter, copper gasket, and manual |
| Optional accessories | Protective rubber cover and carrying bag |
| Optional certificates | KELLER calibration certificate with 5 or 11 measurement points |


Other products