First-class pyranometer RSG1

Budget-friendly price Request a quote
µV, 0÷2Vdc, 4÷20mA, RS485/Modbus | 0 ÷ 2000 W/m² | 0.3 ÷ 3 µm | Class B (ISO 9060) | 255 x 162mm | IP67

El NESA RSG1 es un piranómetro de termopila de primera clase (clase B según ISO 9060) para la medida de la radiación solar global entre 0,3 y 3 µm con respuesta en coseno. Conforme a WMO para la monitorización ambiental y apto para aplicaciones fotovoltaicas (ISO 9060 e ISO/IEC 17025).

Measurement range from 0 to 2000 W/m², typical sensitivity of 10 µV/(W/m²), zero offset < 14 W/m², response time < 20 seconds, expected daily uncertainty < 5%. Double quartz glass dome (K5), IP67 aluminum housing, and natural output in µV, 0÷2Vdc, 4÷20mA or RS485 / Modbus.

Key points

  • First class (class B) according to ISO 9060; compliant with WMO
  • 0 ÷ 2000 W/m², 0.3 ÷ 3 µm, 10 µV/(W/m²)
  • Zero offset < 14 W/m² (at 200 W/m²), stability < ±1.5% per year, non-linearity < ±1%
  • Outputs µV, 0÷2Vdc, 4÷20mA or RS485 / Modbus (ID 003)
  • Anodized aluminum, IP67, -40 ÷ +80 °C, < 630 g

Technical Frequently Asked Questions

What outputs does the RSG1 pyranometer have?

RSG1-N: natural thermopile, 10 µV/(W/m²). RSG1-A: 0÷2Vdc = 0…2000 W/m². RSG1-B: 4÷20mA = 0…2000 W/m², not powered by the loop. RSG1-C: RS485 / Modbus RTU (9600 baud, N, 8, 1; function 0x03; registers 40002 – 40003; standard ID 003).

What makes the RSG1 different?

It is first class: zero offset < 14 W/m², stability < ±1.5% per year, non-linearity < ±1% and expected daily uncertainty < 5%, with double quartz glass dome (K5).

How do I order it?

Sensor (RSG1), output (A, B, C or N), sensor-datalogger cable (05, 10 or xx, length in meters to be specified) and support (SS1 of l = 500 mm, SS2 of l = 900 mm or SS3 of l = 1500 mm). Manufacturer's example: RSG1 A 10 SS2.

Properties

FeatureSpecification
Measurement and operating principle
Measured magnitudeRadiación solar global (respuesta en coseno)
Measurement range0 ÷ 2000 W/m²
Spectral range0.3 µm ÷ 3 µm (AM1.5G)
Typical sensitivity10 µV/(W/m²) over 2π sr
Typical output constant10 mV/(kW/m²)
Resolution< 4 W/m²
Response time< 20 s
Zero offset< 14 W/m² (at 200 W/m²); < ±3 W/m² (ΔT = 5 K/h)
Long-term stability (1 year)< ±1.5%
Cosine response< ±16 W/m²
Temperature response (ΔT 50 K)< 4%
Nonlinearity< ±1 %
Response to tilt (0 – 90°)< ±2%
Expected daily uncertainty< 5%
TransducerThermopyle
ClassificationFirst class (class B according to ISO 9060)
Regulatory complianceISO 9060; WMO; CE
Output signals and interface
Output (version N)10 µV/(W/m²), direct from thermopile
Output (version A)0÷2Vdc = 0…2000 W/m²
Output (version B)4÷20mA = 0…2000 W/m², not powered by the loop
Output (version C)RS485 / Modbus RTU (Half duplex); Standard ID 003; 9600 baud, N, 8, 1; Function 0x03; Registers 40002 – 40003 (IEEE754 32 bits)
Output impedance< 40 Ω
CalibrationBy comparison with a primary standard instrument of superior class and known constant
Power supply, environment, and mechanical data
Food and consumptionSelf-powered in the N version; 10 ÷ 30Vdc < 0.1 W in the others
ProtectionsProtection against short circuits and lightning strikes
Operating conditions-40 to +80 °C
Degree of protectionIP67
MaterialAnodized aluminum and stainless steel screws
Dimensions255 x 162 mm (see drawing)
Weight< 630 g
ReferenceConfiguration
RSG1-ARSG1AModel: RSG1 · Output signal: 0÷2Vdc · Supplied with: Calibration report
RSG1-BRSG1BModel: RSG1 · Output signal: 4÷20mA · Supplied with: Calibration report
RSG1-CRSG1CModel: RSG1 · Output signal: RS485 / Modbus (Standard ID 003) · Supplied with: Calibration report
RSG1-NRSG1NModel: RSG1 · Output signal: Natural in µV (10 µV/(W/m²)) · Supply: With calibration report
NESA RSG1 First Class (Class B) Pyranometer
Dimensions and connections of the NESA RSG1 pyranometer

Other products

Monitran-Logo
Triaxial (X, Y, Z axes) | 3 isolated AC outputs of 100 mV/g ±10% | 2 Hz to 9 kHz ±5% | ±80 g | -25 to 125 °C | 4-pin M12 connector | Aluminum, IP54
DPF Sensors logo
SIL Pressure. Liquids.
NCTE torque sensors.
NCTE Rotary Torque Sensor | 2300 Series | 100Nm | 10,000 RPM
Official logo of ELCIS Encoder, Italian manufacturer of incremental and absolute encoders and industrial measurement systems.
Hollow shaft Ø16 H7 | Housing Ø98 mm | Self-compensating coupling | 1 to 2540 imp/revolution | NPN, PNP, push-pull, driver, TTL | Power supply 5–30 V | IP64 shaft (IP66 with Z option)
LVDT Manufacturer Logo Schreiber Messtechnik
LVDT inductive sensor | Strokes 4 and 8 mm | Housing Ø10 mm | Linearity 0.25 % / 0.5 % | IP68 documented | Moving core or spring-loaded probe | External electronics
Disibeint Electronic, manufacturer of level sensors and control relays
CONDUCTIVE ELECTRODES WITH BUILT-IN AMPLIFIER
Official logo of Novotechnik, manufacturer of linear sensors, position transducers, and industrial encoders
CANopen | 41.5 x 59 mm max. | 14-bit resolution | Shaftless: magnetic placeholder | 12/24 VDC (8 … 34 VDC) | IP67 / IP68 / IP69K | 1 or 2 channels
Official logo of ELCIS Encoder, Italian manufacturer of incremental and absolute encoders and industrial measurement systems.
Absolute, wire Ø 1.35 mm | Stroke 40, 50 or 60 m | Resolution 0.1 or 1 mm | PROFIBUS-DP Class 2 | 10–30 V DC | IP64 (IP65 connection covers) | Linearity ±0.05 % fs.
Official logo of ELCIS Encoder, Italian manufacturer of incremental and absolute encoders and industrial measurement systems.
Through shaft Ø30–50, housing Ø149 | 1 to 3600 impedances/revolution | NPN, PNP, push-pull, line driver | Power supply 5–30 V | IP65 shaft side | 100 kHz (300 kHz optional) | Internal tangential coupling
Shopping Basket