PT100 PT1000 RTD Replicator with galvanic isolation PTi-1000

RTD Replicator for PT100 and PT1000 | 2, 3 and 4 wires | DIN rail | 10-30 V AC/DC

Having trouble duplicating a PT100 or PT1000 signal? Attempting to split a sensor signal using direct taps or jumpers often leads to instability, reading errors due to impedance loading, and risks of dough loops.

He ITAIA PTi-1000 RTD replicator generates independent real resistances at each output, allowing a single sensor to simultaneously feed a PLC, a SCADA and a recorder with complete galvanic independence and without altering the original signal.

Its design is particularly suitable for pulsed or multiplexed measurement systems, where many standard isolators do not maintain a stable reading against the excitation current of the PLC or acquisition system.

Technological advantages for engineering

  • High precision: Typical error of 0.02 % (≈0.05 °C), suitable for critical processes.
  • Complete galvanic isolation: Separation between input, outputs and power supply for maximum noise immunity.
  • Sensor versatility: Compatible with PT100, PT500, PT1000, Ni100, KTYxx and resistive signals.
  • Security in classified areas: Available in certified versions ATEX / IECEx (Ex ia).

Available in versions DH, SH, XD and XS, with resistive output configurations, analog signals 4-20 mA / 0-10 V and display option OLED for local monitoring.

Thanks to your 16-bit resolution, high EMC immunity and compatibility with connections of 2, 3 and 4 threads, The PTi-1000 is a reliable solution for industrial applications where it is required to distribute an RTD signal without loss of accuracy.

Properties

FunctionGalvanically isolated true resistance RTD replicator/duplicator
TicketsRTD / resistance
Compatible sensorsPT100, PT500, PT1000, Ni100, KTxx
Probe connection2, 3 and 4 threads
Cable compensationYes, in a 2-wire configuration
Resistance output range30 Ω … 9.5 kΩ
Number of exits2 or 4, depending on the version
VersionsDH / SH / XD / XS
Analog outputsAccording to version: 0/4-20 mA and 0-10 V
Typical accuracy0,02 %
Resolution16 bits
Input-output delay< 0.5 s
Pulsed/multiplexed measurementCompatible with pulsed and multiplexed measurement systems
Galvanic isolationGalvanic isolation at inlet, outlets and power supply
ATEX CertificationIntrinsically safe version available Ex ia
Feeding10 … 30 V AC/DC
Consumption< 1.2 W (OLED) / 2.5 W peak
Working temperature-30 … +50 °C
Degree of protectionIP20
MountingDIN rail EN 60715
Dimensions22 x 92 x 69 mm / 22 x 92 x 93 mm, depending on the version
ScreenOLED 64×48, depending on the version
RegulationsCE, EN61010-1, EN61000-3/2,-6/1.3, Ex ia option
PTi-1000 RTD replicator for PT100 and PT1000 sensors with galvanic isolation
RTD replicator connection diagram for PT100 and PT1000 in 2, 3 and 4 wires

Other products

Monitran-Logo
Battery Powered | Analog Output (Amps) | Up to 4 Channels | 10 Hours of Autonomy | BNC or Cable Gland Connectors | Low Battery LED Indicator | ABS Carrying Case | Compatible with IEPE Accelerometers | Rugged Field Design | Low Noise Electronics
2-way universal galvanic isolator/converter. Input: mV (DC), Wirelessly configurable ranges (0/10mV…10V), Automatic 0-4/20mA output (Active/Passive), Power supply: 10…30 VDC
Converter for averaging 4 RTD signals to 4/20mA
Expander of 1 digital signal to 16 digital outputs. Expandable to 32, 48, 64, etc. for DIGIMUX16
3-way isolator: ± 0/10V input, isolated excitation (24V/20mA), isolated output ± 0/10V, isolated 24VDC (20.. 30V) power supply
3-channel encoder isolator-amplifier-doubler-adapter. Up to 1 MHz. (adapts differential signals to single-ended signals)
4 Pt100 sensor controller with 2 alarms per relay
Isolator-converter Input: 0-10 Vdc to Isolated output 0-4/20 mA active, Power supply: 24VDC (20.. 30V) isolated
Isolator for passive pickups. Passive output: 4/20 mA, input: 4/20 mA (for passive pickup) – 2 channels
Shopping Basket